background image

Quality in your hands

BUCHI Affiliates:

We are represented by more than 100 distribution partners worldwide. 
Find your local representative at: www.buchi.com

BUCHI Support Centers:

Europe

Switzerland/Austria

BÜCHI Labortechnik AG

CH – 9230 Flawil 
T +41 71 394 63 63
F +41 71 394 65 65
[email protected]
www.buchi.com

Benelux

BÜCHI Labortechnik GmbH

Branch Offi ce Benelux 
NL – 3342 GT Hendrik-Ido-Ambacht 
T +31 78 684 94 29
F +31 78 684 94 30 
[email protected]
www.buchi.be

France

BUCHI Sarl

FR – 94656 Rungis Cedex
T +33 1 56 70 62 50
F +33 1 46 86 00 31
[email protected]
www.buchi.fr 

Germany

BÜCHI Labortechnik GmbH

DE – 45127 Essen
T +800 414 0 414 0 (Toll Free)
T +49 201 747 49 0
F +49 201 747 49 20 
[email protected]
www.buechigmbh.de

Italy

BUCHI Italia s.r.l.

IT – 20010 Cornaredo (MI)
T +39 02 824 50 11
F +39 02 575 12 855
[email protected]
www.buchi.it

Russia

BUCHI Russia/CIS

Russia 127287 Moscow
T +7 495 36 36 495
F +7 495 98 10 520
[email protected]
www.buchi.ru 

United Kingdom

BUCHI UK Ltd.

GB – Oldham OL9 9QL
T +44 161 633 1000
F +44 161 633 1007
[email protected]
www.buchi.co.uk

Germany

BÜCHI NIR-Online

DE – 69190 Walldorf 
T +49 6227 73 26 60
F +49 6227 73 26 70
[email protected]
www.nir-online.de

China

BUCHI China

CN – 200052 Shanghai
T +86 21 6280 3366
F +86 21 5230 8821
[email protected]
www.buchi.com.cn

India

BUCHI India Private Ltd.

IN – Mumbai 400 055 
T +91 22 667 75400
F +91 22 667 18986
[email protected] 
www.buchi.in

Indonesia

PT. BUCHI Indonesia

ID – Tangerang 15321
T +62 21 537 62 16
F +62 21 537 62 17
[email protected]
www.buchi.co.id

Japan

Nihon BUCHI K.K.  

JP – Tokyo 110-0008 
T +81 3 3821 4777
F +81 3 3821 4555
[email protected]
www.nihon-buchi.jp

Korea

BUCHI Korea Inc.

KR – Seoul 153-782
T +82 2 6718 7500
F +82 2 6718 7599
[email protected]
www.buchi.kr

Malaysia

BUCHI Malaysia Sdn. Bhd. 

MY – 47301 Petaling Jaya, 
Selangor
T +60 3 7832 0310
F +60 3 7832 0309
[email protected]
www.buchi.com

Singapore

BUCHI Singapore Pte. Ltd. 

SG – Singapore 609919
T +65 6565 1175
F +65 6566 7047
[email protected]
www.buchi.com

Thailand

BUCHI (Thailand) Ltd. 

TH – Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54
[email protected]
www.buchi.co.th

Asia

America

Brazil

BUCHI Brasil Ltda.

BR – Valinhos SP 13271-200
T +55 19 3849 1201 
F +55 19 3849 2907
[email protected]
www.buchi.com

USA/Canada

BUCHI Corporation

US – New Castle,  DE  19720
T +1 877 692 8244 (Toll Free)
T +1 302 652 3000
F +1 302 652 8777
[email protected]
www.mybuchi.com

South East Asia

BUCHI (Thailand) Ltd.

TH-Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54
[email protected]
www.buchi.com

Middle East

BÜCHI Labortechnik AG

UAE – Dubai 
T +971 4 313 2860
F +971 4 313 2861
[email protected]
www.buchi.com

Latin America

BUCHI Latinoamérica Ltda.

BR – Valinhos SP 13271-200
T +55 19 3849 1201
F +55 19 3849 2907
[email protected]
www.buchi.com

Содержание C-640

Страница 1: ...UV Vis Detector C 640 Operation Manual 11593492C en...

Страница 2: ...es the right to make changes to the manual as deemed necessary in the light of experience espe cially in respect to structure illustrations and technical detail This manual is copyright Information fr...

Страница 3: ...5 Analog range ratio 16 3 6 6 Display brightness 16 3 6 7 Subscription Remote or local control 16 3 6 8 Scanning 17 4 Description of function 18 4 1 Front panel of the instrument 18 4 1 1 LED status 1...

Страница 4: ...5 5 1 Wavelength 28 5 5 2 Time constant 29 5 5 3 Half width 29 5 5 4 Analog range ratio 29 5 5 5 Display brightness 29 5 6 Getting information and scanning 30 5 6 1 General information 30 5 6 2 Inten...

Страница 5: ...l This manual describes the UV VIS Detector C 640 It provides all information required for its safe operation and to maintain it in good working order It is addressed to laboratory personnel in partic...

Страница 6: ...ultaneously and scanning over the entire wavelength range of the detector 200 840 nm becomes possible in conjunction with the BUCHI SepacoreControl software and the control unit C 620 A deuterium lamp...

Страница 7: ...which if not avoided could result in death or serious injury CAUTION Indicates a hazardous situation which if not avoided may result in minor or moderate injury no NOTICE Indicates possible property...

Страница 8: ...at Wear protective goggles Wear protective gloves Additional user information Paragraphs starting with NOTE transport helpful information for working with the device software or its supplementaries NO...

Страница 9: ...vel of hazard Additional warning messages can be found whenever actions and situations described in this manual are related to situational hazards Warning Death or serious injuries by use in explosive...

Страница 10: ...ur during operation of the instrument Legal regulations such as local state and federal laws applying to the instrument must be strictly followed Duty of maintenance and care The operator is responsib...

Страница 11: ...ior of a sample at four selected wavelengths in the range of 200 to 840 nm In conjunction with the SepacoreControl software it is also able to scan the whole range of the spec trum continuously The sc...

Страница 12: ...Each part listed with an order number can be ordered separately as spare part Table 3 2 Standard equipment Product Order no UV VIS detector unit C 640 Preparative flow cell for C 640 11057949 Fitting...

Страница 13: ...6 044357 Test cell for preparative cell Cinch cable 2 pcs For connecting the C 640 directly to the C 660 fraction collector without using the SepacoreControl software 045216 Power cord RS232 cable 3 2...

Страница 14: ...Drift at test cell 254 nm after 1 h 1 x 10 3 AU h Time constant T90 0 5 s 0 75 s 1 s or 2 s Light source Deuterium discharge lamp halogen lamp Maximum flow rate 500 mL min Maximum pressure at the cel...

Страница 15: ...0 nm the lifespan will be reduced To perform the test assemble the black test cell to the cell compartment and measure the light inten sity The result can be interpreted according to the following tab...

Страница 16: ...the noise of the system For these reasons the adaption of the number of processing elements electronic half width of wavelength is optional while the default value HF 4 8 4 nm corresponds to the slit...

Страница 17: ...ossible the sample measurement of the four predefined wavelengths for the four analog outputs is omitted during a scan This means during a scan the measured values on to the analog outputs will not be...

Страница 18: ...nd the graphical display 1 4 2 3 a Power switch ON OFF b LED indicators Lamp green and Error red Push buttons for operating the instrument Graphical display Figure 4 1 Front panel of the instrument 4...

Страница 19: ...mall s in the upper right corner of the display In this mode the keypad of the detector is locked 4 1 3 Push buttons of the front panel Push button Function Right Navigate right or jump to next subpar...

Страница 20: ...right side of the instrument by opening the shutter which at the same time functions as a quick clamping lever for holding the cell in place within the optical path Notice Risk of instrument damage by...

Страница 21: ...h of it In most cases the cell will be set free as soon as the shutter has been opened completely and can be taken away from the cell compartment In case the cell didn t come loose although the shutte...

Страница 22: ...4 3 2 Mounting the cell Enter the cell to the cell compartment slide it onto the groove on the left side of the compartment and hold it on place with two fingers Carefully close the shutter observing...

Страница 23: ...urface Consider the maximum product dimensions and weight We recommend the installation into the chromatography table Obtain the environmental conditions as described in section 3 3 Technical specific...

Страница 24: ...measures such as residual current breakers may be necessary to meet local laws and regulations External connections and extension lines must be provided with an earthed conductor lead 3 pole coupling...

Страница 25: ...d the grey fittings with the grey ferrules 044262 for the 1 16 tubing NOTE The tubing with 1 32 inner diameter together with the corresponding grey fittings can be used for flow rates up to 30 mL min...

Страница 26: ...at Check all connections for leakage after each installation Check all connections for leakage on a regular base during operation Clean the instrument and all accessories thoroughly to remove possibly...

Страница 27: ...remote or subscription mode the keypad of the instrument will be locked signalized by a small key or a small s in the graphical display of the instrument The menu structure of the instrument can be t...

Страница 28: ...r a few seconds After start up is completed and all parameters are restored according to the last settings the idle screen showing the state of the detector and the operating hours of the lamp is disp...

Страница 29: ...this menu the electronic half width of the instru ment can be set See also chapter 3 6 3 Optical and electronical half width spectral resolution 5 5 4 Analog range ratio Within this menu the absorbanc...

Страница 30: ...hardware version 5 6 2 Intensity The Intensity info menu provides information about the current light intensity on all channels See also chapter 3 6 1 Light intensity 5 6 3 Calibration menu To enter t...

Страница 31: ...necessary to scroll through the error description if it does not fit onto the screen all at once If more than one error occurs at once the description text will only reflect the most significant failu...

Страница 32: ...pair work on the instrument Authoriza tion requires a comprehensive technical training and knowledge of possible dangers which might arise when working at the instrument Such training and knowledge ca...

Страница 33: ...s simple cleaning procedure did not solve the problem contact the BUCHI service support or replace the cell windows 6 3 2 Replacing the cell windows If a replacement of cracked or contaminated windows...

Страница 34: ...0 1 0 4 mm 1 1 1 4 mm 2 1 2 4 mm a b c Figure 6 2 Adjustment of the optical path After assembly carefully tighten the screws crosswise using a watchmaker s screwdriver Notice Risk of instrument damag...

Страница 35: ...into operation while the housing is removed or the shutter is open Notice Risk of instrument damage by wrong handling Never touch the glass parts of the lamp with bare fingers Fingerprints on the lamp...

Страница 36: ...and discon nect the quick coupling of the cable the lamp is connected to Replace the lamp by a new one and assemble it follow ing these instructions in reverse order Note Never touch the glass parts...

Страница 37: ...leaning and replace ment of cell windows Warning or error message See next two sub chapters Try resetting the error If resetting does not solve the problem call the service center After switching the...

Страница 38: ...considered as hardware errors Hard ware errors are mainly detected based on physical values that are out of the working range of the instrument In case an error code is displayed and the problem cann...

Страница 39: ...cables and tubings and remove all liquids and dusty residues before packaging the instrument Warning Death or serious poisoning by contact or incorporation of harmful substances Wear safety goggles We...

Страница 40: ...se harmful interference in which case the user will be required to correct the interference at his own expense Fran ais Cet appareil a t test et s est av r conforme aux limites pr vues pour les appare...

Страница 41: ......

Страница 42: ...w buchi com cn India BUCHI India Private Ltd IN Mumbai 400 055 T 91 22 667 75400 F 91 22 667 18986 india buchi com www buchi in Indonesia PT BUCHI Indonesia ID Tangerang 15321 T 62 21 537 62 16 F 62 2...

Отзывы: