5 Putting into operation
30
C-640
Operation Manual, Version C
5 .6
Getting information and scanning
5 .6 .1
General information
The General info menu provides basic information about
the detector. Use the buttons UP and DOWN to navigate
through the different items.
The following information can be retrieved from this
menu:
Serial number, firmware version, firmware date, lamp
hours, available range of wavelength, subscription
frequency, model, Software and hardware version.
5 .6 .2 Intensity
The Intensity info menu provides information about the
current light intensity on all channels.
See also chapter 3.6.1 “Light intensity”.
5 .6 .3
Calibration menu
To enter the Calibration menu press ENTER for 3
seconds.
Press ENTER to set the scan range and to enter scan-
ning.
To reset the accumulated operating hours of the lamp
press ENTER.
Insert the test cell to the cell compartment and press
ZERO. Follow the instructions given on the display.
Set the Wavelength (WL) and press ENTER. Follow the
instructions given on the display.
Press DOWN to perform a standard calibration.
Follow the instructions given on the display.