3 Technical data
14
C-640
Operation Manual, Version C
3 .3
Technical specifications
Table 3-4: Technical specifications
Average power consumption
max. 90 W
Power supply voltage
100–240 VAC ± 10%
Frequency
50 / 60 Hz
Fuse
2.5 A / 250 V
Dimensions (w
×
d
×
h)
135
×
418
×
185 mm
Weight
6
kg
Wavelength range
200–840 nm (256 elements on CCD)
Typical spectral half width
10 nm
Accuracy of adjustment
± 1 nm
Reproducibility
± 0.5 nm
Noise level at test cell
± 5 x 10
-5
AU at 254 nm, TC 0.75 s
Drift at test cell (254 nm after 1 h)
1 x 10
-3
AU/h
Time constant (T90)
0.5 s ; 0.75 s; 1 s or 2 s
Light source
Deuterium discharge lamp / halogen lamp
Maximum flow rate
500 mL/min
Maximum pressure at the cell
20 bar / 290 psi
Output for integrator
1 V/AU (in digital form only)
Interface
RS232
Installation category
II
Degree of protection
IP20
Pollution degree
2
Environmental conditions:
Temperature
5–40 °C
for indoor use only
Altitude
up to 2000 m
Humidity
max. relative humidity 80% for temperatures up to
31 °C, linearly decreasing to 50% at 40 °C
Table 3-5: Material and Approvals
Materials in contact with mobile phase
PTFE / fused silica / stainless steel
Approvals
CE, CSA