46
Melting point
B-540
9 Appendix
Dimensions of the unit (width x height x depth)
170 x 210 x 340 mm
Dimensions of packaging (width x height x depth)
340 x 390 x 580 mm
Weight
5.4 kg
Temperature range
25 °C - 400°C
Cooling-down rate 350°C – 50°C
max. 10 min
Heating-up speed 50°C – 350° C
max. 7 min
Measuring accuracy up to 100°C
+/- 0.3°C
Measuring accuracy up to 250°C
+/- 0.5°C
Measuring accuracy up to 400°C
+/- 0.8°C
Temperature ramps
0.1/0.2/0.5/1.0/2.0/5.0/10.0/20.0°C min.
Mains voltage
100 – 120 V/ 50 or 60 Hz
220 – 230 V/50 or 60 Hz
Power consumption
approx. 200 W
Current consumption
approx. 2 A (100-120 V) / 1 A (220-230 V)
Data saving with current failure
The status of the unit is stored
Restarting is required
Ambient temperature
5 – 40°C
Table 6: Technical data
Fault
Cause
Remedy
STORED DATA LOST
The battery is exhausted
Press the Menu key for 5 seconds.
PLEASE CHECK BATTERIE
Order new battery and have it installed by techn.
AND SET TIME AND DATE
service. Date and time to be reprogrammed.
CALIBRATION INTERVAL
The selected calibration
Press the Menu key for 5 seconds.
EXCEEDED!
time has expired.
Calibrate the unit.
LAST CAL: XX:XX
(See chapter 6, Maintenance)
The Citizen printer does
Configuration
The SEL key on the printer must
not print
be actuated whenever he printer is switched on.
CALLSERVICE
Damage in the unit
Switch off the unit (main switch).
ERROR 001/099
Press START-key.
Switch on the unit and keep holding the START-key.
Wrong melting points.
Wrong transmission.
Always remove capillaries when
switching the unit on.
Calibration
Calibrate the unit.
Table 5: Error messages
9.7
Technical data
9.6
Error messages