45
Melting point
B-540
9 Appendix
9.5
Print-Out
What
Input/Change
Company name
Configuration menu 2
Date of determination
Configuration menu 2
Method name
Main menu 2
Identification of the determination
Main menu 2
Start temperature of the determination
Main menu 1
Heating rate selected during the determination
Main menu 1
End temperature of the determination
Main menu 1
Determined melting points per capillary
Signature of the controller
Date of the last calibration
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
9.4.5 Error messages
Memory is empty - no entries.
Program name exists already. Insert a new one.
Program not stored, no name has been inserted.
Memory full. Clear programs to obtain free memory space.
No printer selected. (Configuration menu)
Check the battery (Lithium cell in the unit) and set new time and date.
Calibration interval exceeded.
(Keep pressing MENU-key to go on working)
Last calibration.
Switch off the unit (main switch).
Keep pressing the START-key and switch on the unit.
Call customer service (see chapter 6.4).
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Figure 14: Print-Out
TEMPERATURE
: XXX.X°C
MEMORY EMPTY !
PRESS ENTER
TEMPERATURE
: XXX.X°C
NAME EXIST ALREADY
PRESS ENTER
TEMPERATURE
: XXX.X°C
NO NAME - NOT STORED!
PRESS ENTER
TEMPERATURE
:
XXX.X°
MEMORY FULL !
PRESS ENTER
TEMPERATURE : XXX.X°C
NO PRINTER SELECTED
PRESS ENTER
STORED DATA LOST
PLEASE CHECK BATTERY
AND SET TIME AND DATE
CALIBRATION INTERVAL
EXCEEDED!
LAST CAL.: XX.XX.XX
CALL SERVICE
ERROR: 001/099