8 | Help with faults
BÜCHI Labortechnik AG
32/38
Operation Manual Inert Loop B-295 SE
u
Disconnect the sensor cable by turning the
plug.
STOP
START
u
Remove the oxygen sensor.
8.2.2
Installing the oxygen sensor
Precondition:
R
The oxygen sensor is removed. See Chap-
ter 8.2.1 "Removing the oxygen sensor",
page 31
u
Remove the foil from the oxygen sensor.
u
Install the new oxygen sensor.
u
Reconnect the sensor cable.
u
Calibrate the sensor. See Chapter 8.1 "Cali-
brating the oxygen analyzer", page 29