BÜCHI Labortechnik AG
Cleaning and servicing | 7
Operation Manual Inert Loop B-295 SE
25/38
7
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec-
tion.
Any servicing and repair work which involves opening up the casing may only be car-
ried out by BUCHI service technicians.
u
Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
7.1
Regular maintenance work
Component
Action
Frequency
Checking seals
u
Check for damages and wear
u
Replace the seals if necessary
Weekly
Quick connectors
u
Check the seals for damages and
wear
u
Replace the seals if necessary
Weekly
Solvent recovery
vessel
u
Check the solvent recovery vessel for
defects
u
If defective, replace the solvent recov-
ery vessel
u
Check the seals for damages and
wear
u
Replace the seals if necessary
Weekly
Casing
u
Wipe down the casing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Ventilation slots
u
Remove dust and foreign objects
from the ventilation slots using com-
pressed air or a vacuum cleaner
Monthly
Oxygen Sensor
u
Calibrate the oxygen analyzer. See
Chapter 8.1 "Calibrating the oxygen
analyzer", page 29
Yearly
7.2
Cleaning the instrument
u
Disconnect the power cable.