BÜCHI Labortechnik AG
Installation | 5
Operation Manual Inert Loop B-295 SE
19/38
NOTICE
Risk of property damage and diminished performance due to use of un-
suitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the re-
quirements of the device. If other power cables that do not meet those requirements
are used, the device may be damaged and/or its performance diminished.
u
Use only the power supply cables supplied with the product or ordered separately
from BUCHI.
u
If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.
u
Make sure that all connected devices are earthed.
u
Plug the power cable into the connection on the instrument. See Chapter 3.2
"Configuration", page 11
u
Plug the mains plug into the mains outlet socket.
5.4
Installing the oxygen sensor
u
Open the front cover oxygen sensor.
u
Remove the grounding cable from the front
cover oxygen sensor.
u
Remove the screw plug.