3 | Product description
BÜCHI Labortechnik AG
14/38
Operation Manual Inert Loop B-295 SE
Inert Loop
B-295 SE
230 V / 50
Hz
Inert Loop
B-295 SE
220 V / 60
Hz
Inert Loop
B-295 SE
200 V / 50
Hz
Inert Loop
B-295 SE
200 V / 60
Hz
Overvoltage category
II
II
II
II
Degree of protection
l
l
l
l
Pollution degree
ll
ll
ll
ll
Dimensions (W x D x H)
657 x 653
x 680 mm
657 x 653
x 680 mm
657 x 653
x 680 mm
657 x 653
x 680 mm
Weight
98 kg
98 kg
99 kg
99 kg
Refrigerant
R449A,
CFC free
R449A,
CFC free
R449A,
CFC free
R449A,
CFC free
Filling amount of refrigerant
0.98 kg
0.98 kg
0.98 kg
0.98 kg
CO
₂
-equivalent
1369.06 kg
CO
₂
1369.06 kg
CO
₂
1369.06 kg
CO
₂
1369.06 kg
CO
₂
Safety Group Refrigerants
(ASHRAE)
A1 (lower
toxicity, no
flame
propaga-
tion)
A1 (lower
toxicity, no
flame
propaga-
tion)
A1 (lower
toxicity, no
flame
propaga-
tion)
A1 (lower
toxicity, no
flame
propaga-
tion)
Maximum allowable pressure
(PS) [low pressure side]
22 bar
22 bar
22 bar
22 bar
Maximum allowable pressure
(PS) [high pressure side]
26 bar
26 bar
26 bar
26 bar
Cooling capacity at the process
gas (at -23 ° C)
800 W
920 W
800 W
920 W
Compressor cooling capacity
(according to EN12900, at -15 °
C)
1334 W
1925 W
1334W
1925 W
Control range process gas tem-
perature
min -30 to
8°C
min -30 to
8 °C
min -30 to
8 °C
min -30 to
8 °C
Minimum distance to all sides
300 mm
300 mm
300 mm
300 mm
EMC to EN 61326-1 Emmission
Class B
Class B
Class B
Class B
EMC to EN 61326-1 Immision
Industry
Industry
Industry
Industry
3.4.2
Ambient conditions
For indoor use only.
Max. altitude above sea level
2000 m
Ambient temperature
5‒40 °C
Maximum relative humidity
80 % for temperatures up to 31 °C
decreasing linearly to 50 % relative hu-
midity at 40 °C
Storage temperature
max. 40 °C
Содержание B-295 SE
Страница 1: ...Inert Loop B 295 SE Operation Manual STOP START 11594141 B en...
Страница 37: ......