Mini Spray Dryer

 B-290

Quick Operation Guide 

Switch  

Aspirator on

Set Inlet temperature,

Switch Heater on

Open  

compressed air

Distilled water ready, 

start peristaltic pump

Reduce inlet  

temp. and/or  

increase  

pump rate

Increase inlet  

temp. and/or  

reduce 

pump rate

Adjust Nozzle  

cleaning

Change from distilled 

water to product

Change from product 

to distilled water

Inlet Temp. reached

Spray Drying in process

Outlet Temp.
too high

Wet glass

product sticking

Generates  the  gas  stream  in  the  system.  100%  re-
commended,  so  efficient  drying  and  best  separation 
is obtained.

For aqueous solutions, inlet temperatures are between 
100 and 220°C. Lower range for heat sensitive product.  
Typical set point 150°C.

By opening the needle valve on the flow control, the 
compressed air for the feed dispersion is set. Typical 
level is 4 cm (approx. 600 l/h).

Close  peristaltic  pump  bed.  The  process  is  started 
with pure water to get the drying conditions . The outlet 
temp. will decrease due to the consumption of energy 
by the evaporation.
Typical set point of pump 30%.

The outlet temp. defines the maximal thermal load for 
the product. Adjust inlet temp. and pump rate to obtain 
a certain outlet temp. Lowering the aspirator rate will 
also decrease the outlet temp.

In case of suspensions or product which tend to clog 
the nozzle, the nozzle cleaning is switched on. Higher 
number (2-9) means more cleaning pushes.

The spray drying starts as soon as the tube is switched 
from  pure  water  to  product  (by  feed  switch  valve  or 
manually). The outlet temp. will slightly increase becau-
se less water has to be evaporated, eventually adjust 
inlet temp.

Switch back to pure water to clean tube and nozzle. 
Additional cleaning effect by drawing air bubbles.