
Description
2
CLS Pro 600 – Line and Contrast Sensor
13/108
③
Display Area
The content shown in this area depends on the mode set-
ting. Detailed descriptions can be found in Section 6.2.1 for
the setup mode and in Section 6.2.2 for the scanning mode.
2.4.2
Set Position
The set position is the required line / contrast transfer point posi-
tion in the sensor scanning area. It can be set up in
Automatic
mode using the control panel’s arrow keys
(see Chapter 6.3.2)
.
The set point position will be indicated by a
white triangle
dis-
played on the control panel’s screen.
Fig. 8: SET POINT Position without deviation
①
Center of Measuring Range
②
Contrast Quality (graphical)
③
Standard SET Position
The deviation will be faded in as plaintext during displacement of
the set position. The deviation can take positive values and negat-
ive values as well.