9
1.
PF
(Total Power Factor) is displayed automatically in the secondary mini display.
2. Annunciator “
A-lags-V
” turns on to indicate an
inductive
circuit is being measured.
That is, the Current waveform is lagging the Voltage waveform, and the phase-shift
angle
盈
is “+”.
3. On the contrary, together with significant
PF
values,
WITHOUT
“
A-lags-V
” being
turned on indicates that a
capacitive
circuit is being measured. That is, the Current
waveform is leading the Voltage waveform, and the phase-shift angle
盈
is “-”.
Note:
1. Under proper measurement setups for load circuits, the
W
(real power) readings are
always positive. Negative
W
readings indicate reversed clamp-on jaws direction or test
leads polarities, or even incorrect voltage lines are being measured as in 3-phase
measurement setups. Correct them for proper “
A-lags-V
” indications.
2. When encountering largely distorted waveforms, “
A-lags-V
” detection might be
affected due to the influence of harmonics. It is recommended to manage (e.g. filter
out) harmonics problems before measuring/dealing with phase-shift problems.