32
Z31900_00_01
Technical Data
Pneumatic
i
Pressure supply above 7 bar will be automatically repressed to avoid damage to the
SampleXpress.
4.3
Operating Conditions
Environment
For the appropriate temperature see also the Bruker site planning guides on the BASH
CD (Bruker Advanced Service Handbook):
Data
Value
Unit
Operating pressure
4-7
bar
Compressed air requirement, minimum
> 100
l/min.
Oiling, maximum
0.005
mg/m
3
Table 4.3 Pneumatic Connection Values
Data
Value
Unit
Temperature range
5-30
°C
Relative humidity at 31 °C, maximum
< 80
%
Decreasing linear till relative humidity < 50% at 40
°C, maximum
< 50
%
Table 4.4 Operating Environment
Manual
Bruker Part Number
Site Planning for AVANCE Systems 300-700 MHz (UM)
Z31276
Site Planning for AVANCE Systems 750 -950 MHz (UM)
Z31686
Table 4.5 Bruker Site Planning Guides
Содержание SampleXpress
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy SampleXpress 001 Sample Changer User Manual ...
Страница 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Страница 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Страница 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Страница 104: ...104 Z31900_00_01 Maintenance ...
Страница 116: ...116 Z31900_00_01 Error Codes ...
Страница 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Страница 122: ...122 Z31900_00_01 Contact ...
Страница 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Страница 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Страница 158: ......
Страница 160: ...160 Z31900_00_01 Warning Signs ...
Страница 164: ...164 Z31900_00_01 Figures ...
Страница 166: ...166 Z31900_00_01 Tables ...
Страница 168: ...168 Z31900_00_01 Glossary ...
Страница 172: ...172 Z31900_00_01 Index ...
Страница 173: ......