73
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.4
Cooling down from 300 K to 80 K
Preconditions
•
Magnet Stand mounted (refer to the manual of the Magnet Stand).
•
Helium flow system mounted (see chapter
”Mounting the Helium Flow System”
).
•
Cryostat is evacuated with a vacuum less than 5 x 10
-8
bar.
•
Cryogenic Refrigerator mounted completely.
•
Transportation dewar with sufficient dry and warm nitrogen gas.
•
Transportation dewar with sufficient warm helium gas.
•
ACD (Automatic Cooling Device) mounted on a transportation dewar with sufficient
liquid nitrogen.
WARNING
Cryogenic Agents
)
Gas under Pressure
(see
Figure 4.77: Assembly for cooling down from 300 K to 80 K with ACD
1 Current lead turret
2 Helium fill-in turret
3 ACD (Automatic Cooling
Device)
4 Transportation dewar with
liquid nitrogen
5 Transportation dewar with
warm nitrogen gas
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...