71
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.3.7
Mounting the Temperature Control
The temperature of the refrigerated radiation shield inside the cryostat during usual ope-
ration must be controlled and in case of failure the operator will receive an alarm.
Set the alarm temperature level to 90 K at MICS and at the External Temperature
Control Unit.
The alarm will be given in four different ways:
•
Alarm message at the computer of the magnet system via MICS.
•
Alarm message by email generated via MICS.
•
Alarm message at the display of the External Temperature Control Unit.
•
Audible alarm from the External Temperature Control Unit.
The cryostat has two PT 100 temperature sensors for monitoring. One sensor is con-
nected to the BSMS board of the console with supplied VTA MAG-RS unit and Cable Set
BSVT, controlled with MICS. The other sensor is connected to the External Temperature
Control Unit with the supplied cable.
i
The two PT 100 sensors should be used in a parallel manner to prevent any lack of
monitoring due to failure of the sensors or the power supply.
Thus: connect the console to another electrical circuit than the External Temperature
Control Unit.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...