57
ZTKS0156 / Z31820 / Rev.: 03
Assembling
i
The spikes of the base plate should be protected with the supplied plastic cap nuts until
the rotary valve column is in its final position to prevent damage to the floor surface.
These plastic cap nuts have to be removed before operating the magnet system.
Figure 4.39: Mounting the Rotary Valve Column - step 1
1.
Put the base plate (1) upside down.
2.
Arrange the three base supports (2) as
shown in the figure with the tapered bores
(3) on topside.
Figure 4.40: Mounting the Rotary Valve Column - step 2
3.
Mount the three base supports to the base
plate with three M 8 x 25 screws (1) per sup-
port.
4.
Tighten the screws (torque 20 Nm).
Figure 4.41: Mounting the Rotary Valve Column - step 3
5.
Attach the spikes (1) into the middle thread
of the three base supports.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...