ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.2.2
Mounting the Vacuum Valve at Evacuated Cryostat
i
The sealing plug can not be moved until a correct vacuum was applied at the KF 40
flange at the Vacuum Valve.
Figure 4.28: Mounting the Vacuum Valve at evacuated cryostat
1.
Remove the protection cap of the vacuum
flange at the bottom plate (2) of the cryostat.
2.
Install the valve stem (10) onto the sealing
plug (3) and tighten it slightly.
3.
Turn the Vacuum Valve in the desired posi-
tion (KF 40 flange (9) looking outwards).
4.
Place the two half rings (7) as shown into
the slots of the valve body (1).
5.
Fix the half rings (7) with four M 6 x 12 (8)
screws.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...