116
ZTKS0156 / Z31820 / Rev.: 03
Maintenance
7.4
Maintenance Work at the Cryogenic Refrigerator
Approved Persons: Bruker Service only
7.4.1
Service Kit
For work at the Cryogenic Refrigerator use the Service Kit (Material No. Z119869).
Figure 7.1:
Service Kit
The Service Kit includes:
•
Non-ferromagnetic wrench set
•
Aeroquip® adapters
•
Helium Manifold
•
Flex Line for servicing
•
Frequently needed spare parts
Figure 7.2:
Tool Tray
Optionally: Tool tray
The Tool Tray fits on each of the RS towers. It is
recommended for depositing tools and parts
necessary for work on top of the cryostat to avoid
falling objects.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...