107
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
6.3
Troubleshooting Work
6.3.1
After a Quench
Quench while in persistent mode:
1.
Wait until the quench valves are closed and no helium evaporate out of the quench
valves.
2.
Wait until there is no helium vapor visible in the room or the ceiling to make sure
there is sufficient oxygen in the room.
3.
Switch off the alarm at the External Temperature Control Unit.
4.
Check the globes in the quench valves for their correct position.
5.
Remove probe and shim system to prevent icing from the shim system.
6.
Check the nitrogen turrets for icing.
7.
Start the refill with liquid helium as soon as possible after the quench (within one
hour after the quench; refer to the supplied refill manual).
8.
Contact Bruker Service immediately.
Figure 6.1:
Quench picture
A quench is the very fast de-energizing of
the magnet by loss of its superconductivity.
The stored magnetic energy is converted
into heat, which promotes rapid evapora-
tion of large quantities of helium.
After an appearance of a quench contact
Bruker Service immediately.
WARNING
Cryogenic Agents
Quench
(see
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...