59
Bruker Optik GmbH
IFS 125M User Manual
Operation 5
3
Unscrew the Teflon-made counter nuts from
the 2 detector hoods.
i
The Teflon-made counter nuts are only
available if you use a liquid nitrogen-
cooled detector.
4
• Use a cross-head screwdriver (size: PH2)
and rotate the 4 screws of the top lid coun-
terclockwise.
➣
You have to rotate over the snap-in resis-
tance. Then, the screw is pushed
upwards, and you can loosen it.
• Carefully lift off the sample compartment
cover.
5
Use a flat-tip screwdriver to loosen the two
knurled thumb screws, located on the bottom
of the trough.
6
Completely remove the trough out of the sam-
ple compartment.
7
Repeat step 1 to remove the cover of the detector and interferometer block.
Table 5.9: Changing detector - Removing covers
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...