38
IFS 125M User Manual
Bruker Optik GmbH
Description 4
4.7
Input ports
Ensuring electromagnetic shielding requires a conductive connection between acces-
sory and spectrometer, and a complete metallic enclosure of the accessory without any
slits or openings larger than 4 cm are required. Unused ports have always to be covered
with (metallic) blank flanges.
If the port opening has a clear aperture of more than 4 cm, the port will not be shielded
any more according to electromagnetic compatibility as soon as the blank flange is
removed.
Figure 4.7:
Input ports
NOTE
Clear port aperture of port opening >4 cm:
No shield safety according to EMC
a
requirements.
➣
Accessory connected must ensure complete EMC shielding.
a. EMC: Electromagnetic Compatibility
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...