Installation 3
Power supply
a
:
The spectrometer power supply unit automatically
adapts to the most common power sources.
• Valid voltage range: 100 V AC to 240 V AC
• Valid frequency range: 50 to 60 Hz
• Additionally, any power supply via any 24 V DC
source is possible (load 5 A)
No uninterruptible power supply:
Property damage.
➣
Always ensure uninterruptible power supply (UPS)
and/or a line conditioner.
How to interrupt power
supply:
• Either switch off the spectrometer using the ON/OFF
switch, which is located on the front side of the
mobile electronics unit or
• unplug the power cable from the primary power
receptacle (AC inlet), which is located at the power
supply unit or
• unplug the power cable from the mains outlet
Space:
• Stable and horizontal base which is able to bear the
weight of at least 160 kg
• For exact dimensions refer to appendix A.
Temperature variations:
• Less than 1°C per hour and not more than 4°C per
day
• Avoid temperature variations as they can impair the
results of long-term measurements
Vibration:
Do not locate the spectrometer in high traffic areas,
nor near vibration sources (ventilation hoods, air con-
ditioners etc.).
a. If you connect external accessories to the spectrometer interface such as R232 (serial), Ethernet or
LPT1 (parallel), these accessories must comply with the requirements outlined in EN 61010 or EN
60950. This means that you always have to connect a SELV circuit (Safety Extra Low Voltage Circuit) to
a corresponding SELV circuit and not to any other circuit type.
Table 3.2: Site requirements
NOTE
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...