Troubleshooting 7
7.4
Spectrometer problems indicated by file parameters in
OPUS
To have the spectrum file parameters displayed in OPUS proceed as follows:
1.
Right click the spectrum file in the OPUS browser.
2.
Select
Show Parameters
from the pop-up menu.
➣
The following dialog opens:
3.
Select
Instrument Parameters
and
Data Acquisition Status
.
➣
Any value unequal to zero indicates an error during data acquisition. Bit coded
value, bit 13 set (= 8192 decimal) means Buffer overflow in the optics controller.
This indicates a severe problem in the data transfer from the DSP of the optics to
OPUS. Data has been lost and the resulting file is corrupt. This can be either due
to an unreliable network connection, or to a too high a data rate. In the latter
case, the scanner velocity should be reduced.
Figure 7.3:
Spectrum file parameters displayed in OPUS
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...