42
ZTKS0213 / Z31939 / Rev.: 01
Operation
5.6.3
Energizing the Magnet
Precondition
•
Magnet system at base temperature (less than 3.5 K).
•
All green indicators on the front panel of the BCFC "ON".
•
Safety check passed.
Procedure
Energize the magnet according to section
"Energizing and Deenergizing in Main Mode"
. Watch the field indicator. The indicator must become active (ON) before the
magnet reaches 1 T (10 Amps). If the indicator remains inactive (OFF), refer to the
troubleshooting section of this manual (see
).
Figure 5.16: Energizing the magnet and checking the indicator
Do not operate the magnet system with a defective magnetic field indicator.