ZTKS0213 / Z31939 / Rev.: 01
Operation
5.2
Evacuating the Cryostat
5.2.1
General View
The vacuum valve (Material No. Z53420) is necessary to evacuate the vacuum chamber
of the cryostat after assembling the magnet system.
A turbomolecular pump (pumping speed
≥
80 l/s N
2
) is required to evacuate the
cryostat.
Do not use a diffusion pump.
Figure 5.1:
Vacuum valve assembly
1 Valve operator body
2 KF 40 flange for the pumping line
3 Two half rings with Allen screws M6 x 12
4 Valve stem
Tilting of the vacuum valve damages the O-ring of the vacuum valve. Thus, pay
attention to the orientation of the axis during insertion of the vacuum valve.