71
Z31823_00_05
Troubleshooting
11.1.1.6
Limit Data Helium Level Failed
It was not possible to read the limit data for the helium level from the EEProm, thus
supervision of the helium level is not possible.
Solution
: Reload the values from the backup database to the BMPC/2 as shown for
Limit Data P1 fail.
11.1.1.7
Calibration T1 Failed
It was not possible to read the calibration values for the calculation of the magnet tem-
perature T1 from the EEProm, thus calculation and supervision of the temperature T1 is
not possible.
Solution:
A software reset of the BMPC/2 Control Unit usually helps (see also
). If not, you will need to recalibrate the temperature T1 (see also
"Calibrate Temperature Linearization Function" on page 68
).
11.1.1.8
Calibration T2 Failed
It was not possible to read the calibration values for the calculation of the magnet tem-
perature T2 from the EEProm, thus calculation and supervision of the temperature T2 is
not possible.
Solution:
A software reset of the BMPC/2 Control Unit usually helps (see also
). If not, you will need to recalibrate the temperature T2 (see also
"Calibrate Temperature Linearization Function" on page 68
).
11.1.1.9
Calibration T5 Failed
It was not possible to read the calibration values for the calculation of the pump cabinet
temperature T5 from the EEProm, thus calculation and supervision of the temperature
T5 is not possible.
Solution:
A software reset of the BMPC/2 Control Unit usually helps (see also
11.1.1.10
Heater Initialization Failed
The initialization values for the pressure regulation for pressure P2 using the magnet
heater could not be read from the EEProm, thus the pressure regulation could not be
activated.
Solution:
Reenter the data for the pressure regulation using the menu
Tools - Options -
Heater & APD Setting
.
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......