Planning details
Specific solutions are possible in special cases for achieving load distribution in existing build-
ings. These solutions should be analyzed in relation to their vibration behavior. Discuss the sit-
uation and possible counter-measures with the planners from your local Bruker BioSpin MRI
office.
6.4.2
Magnet foundation
The ferromagnetic density should not exceed 100kg/m
2
in the area of the magnet foundation. If
this limit cannot be maintained for new buildings, non-magnetic reinforcement materials (e.g.
high-yield austenitic steels) should be considered.
The magnet foundation must be constructed in the form of a bonded screed. If an HF shielded
room is used, the bonded screed must extend at least over the surface area of the box.
Construction of the magnet foundation, if the original floor construction at the magnet installation site is based on a floating floor.
Figure 6.11: Construction of the magnet foundation, if the original floor construction at the magnet installa-
tion site is based on a floating floor.
1
accessible flooring, constructed as a floating floor
2
absorption for sound impact decoupling
3
base plate of the building
4
magnet foundation, constructed as bonded screed
Содержание BioSpec 117 AVANCE III
Страница 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Страница 6: ...Contents 6 ID 182463115...
Страница 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Страница 72: ...72 ID 182463115 Planning details...
Страница 82: ...82 ID 182463115 Delivery and transport...
Страница 84: ...84 ID 182463115 Moving or dismantling the MR system...
Страница 92: ...Figures 92 ID 182463115...
Страница 94: ...Tables 94 ID 182463115...
Страница 101: ...ID 182463115 107 Lastpage...
Страница 102: ...Bruker Corporation info bruker com www bruker com...