54
ZTKS0168 / Z31960 / 05
Troubleshooting
6.3
Troubleshooting Work
6.3.1
After a Quench
Quench while in persistent mode:
1.
Wait until helium stops evaporating and the quench valves are closed.
2.
Wait until there is no helium vapor visible anywhere to make sure there is sufficient
oxygen in the room.
3.
Check that the globes in the quench valves are in the correct position.
4.
Remove probe and shim system to prevent formation of ice on the shim system.
5.
Start the refill with liquid helium as soon as possible.
6.
Contact Bruker Service immediately.
Figure 6.1: Magnet system during a quench
A quench is the very fast de-energizing of
the magnet by loss of its superconductivity.
The stored magnetic energy is converted
into heat, which promotes rapid evaporation
of large quantities of helium.
If a quench occurs contact Bruker Service
immediately.
WARNING
Cryogenic Agents (see
Содержание 300'89 Ascend
Страница 1: ...Version Innovation with Integrity 300 89 Ascend 05 User Manual NMR...
Страница 6: ...6 ZTKS0168 Z31960 05 Table of Contents...
Страница 8: ...8 ZTKS0168 Z31960 05 Contact...
Страница 28: ...28 ZTKS0168 Z31960 05 Safety...
Страница 34: ...34 ZTKS0168 Z31960 05 Transportation...
Страница 35: ...35 ZTKS0168 Z31960 05 4 Assembling Approved Persons Bruker Service only...
Страница 36: ...36 ZTKS0168 Z31960 05 Assembling...
Страница 57: ...57 ZTKS0168 Z31960 05 8 Disassembling Approved Persons Bruker Service only...
Страница 58: ...58 ZTKS0168 Z31960 05 Disassembling...
Страница 81: ...81 ZTKS0168 Z31960 05 Technical Data MS 300 89 Ascend 9 21 Current Lead Figure 9 12 Current lead 55 pin...
Страница 90: ...90 ZTKS0168 Z31960 05 Technical Data MS 300 89 Ascend...
Страница 92: ...92 ZTKS0168 Z31960 05 Warning Signs...
Страница 94: ...94 ZTKS0168 Z31960 05 List of Figures...
Страница 96: ...96 ZTKS0168 Z31960 05 List of Tables...
Страница 98: ...98 ZTKS0168 Z31960 05 Index...
Страница 100: ...Bruker Corporation info bruker com www bruker com Order No Z31960...