Calibration
Rev. F
Scanning Probe Microscope Training Notebook
63
Figure 21.0a
Baring/Depth Analysis Screen
If the measurement is not 200nm + 1nm, calculate the new sensitivity by the following formula:
Input the new sensitivity into the
Z sensitivity
parameter in the
Microscope
>
Calibrate
>
Z
menu, then repeat this procedure until the scanner is calibrated.
New Sensitivity =
Old Sensitivity x Actual Depth (200nm)
Measured Depth