General Hardware Setup
3. Select the Q value of the insert according to the expected line widths (higher Q
for line widths up to 100 kHz). Please note that multiturn coils, especially multi-
filament coils, have an intrinsically low Q.
Figure 3.33. RF Setup of a Wideline Single Frequency Probe
Shifting the Probe Tuning Range
3.7.2
Most probes cover a fairly wide frequency range. Changing the frequency range
of a probe requires either a change in the inductance or capacitance of the circuit.
The inductance of a circuit is hard to change unless a coil is mechanically length-
ened or shortened. Most probes are tuned over a certain range by variation of a
capacitance. The frequency range is then determined by the minimum and maxi-
mum capacitance that can be set. In order to make the inductance as high as pos-
sible (since the signal from the oscillating magnetization is detected in the
inductive part), one usually chooses a capacitance with very small minimum ca-
pacitance, which again means a not large enough maximum capacitance. So in
most probes additional tuning components must be inserted (removed) to achieve
the full tuning range. The highest signal to noise is always reached with maximum
inductance and minimum capacitance, i.e. at the high end of the tuning frequency
achieved with maximum inductance.
1. Tuning capacitance
2. Exchangeable NMR coil
3. Exchangeable symmetrization and Q-reduction
1
2
3
Содержание Solid State NMR
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy Solid State NMR 002 AVANCE Solids User Manual...
Страница 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Страница 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Страница 104: ...104 327 BRUKER BIOSPIN User Manual Version 002 Practical CP MAS Spectroscopy on Spin 1 2 Nuclei...
Страница 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Страница 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Страница 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Страница 172: ...172 327 BRUKER BIOSPIN User Manual Version 002 SUPER Figure 13 3 The Acquisition Parameter Window eda...
Страница 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Страница 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Страница 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Страница 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Страница 284: ...284 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 1D Figure 22 9 WPMLG CRAMPS After Optimization Digital Acquisition...
Страница 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Страница 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Страница 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Страница 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Страница 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...