Technical Manual Version 002
BRUKER BIOSPIN
37 (55)
Authorised functions
NP - Nitrogen Heater Power
4.8.10
Write
WARNING :
Space characters are not allowed.
Read
NB
: After power on the nitrogen heater power is at «0».
Syntax :
EOT 0 0 0 0 STX
NP
state
ETX BCC
Response :
ACK
Description :
Allows to switch nitrogen heater power ON or OFF.
Rules :
State can be «0» or «1».
«1» switch LN
2
heater ON.
«0» switch LN
2
heater OFF.
Syntax :
EOT 0 0 0 0
NP
ENQ
Response :
STX
NP
State ETX BCC
Description :
Allows to read Nitrogen Power.
Rules :
State can be «0» or «1».
State at «1» means that LN
2
heater is ON.
State at «0» means that LN
2
heater is OFF.