Operating & Service Manual Version 001
BRUKER BIOSPIN
27 (47)
6
Servicing the BLA
6
The BLAH2000-I E C Amplifier 300MHz provides diagnosis and servicing web
pages relies on HTTP, allowing service access with any web browser.
Accessing the BLA Amplifier
6.1
The BLAH2000-I E C Amplifier 300MHz is accessible via the BLA control board
with its IP address.
The IP address is given during "
cf
" by using TOPSPIN 2.xx software on the
workstation.
In case of problems :
•
Check the RJ45 cabling between amplifier, Ethernet switch and workstation.
•
Check the Ethernet switch power.
•
Check if the green LED on the amplifier RJ45 connector lights up.
•
Check the front panel of the amplifier, LED's indi32V, +15V, -15V and
+3.3V ON must have lit.
To access the BLAH2000-I E C Amplifier 300MHz, type "
ha
" in TOPSPIN 2.xx or
better and choose the BLA that should be accessed or start your favourite web
browser and type the given IP address as URL.