background image

9

Compliance with Standards

Specifications – Noise Monitoring Terminal Plus Type 3639-E and Noise Monitoring Terminal Advanced 
Type 3639-G

Specifications apply to all configurations unless otherwise stated

STANDARDS

Conforms with the following National and International Sound Level 
Meter Standards:

IEC 61672-1 (2002-05) Class 1

IEC 60651 (1979) plus Amendment 1 (1993-02) and Amendment 2 
(2000-10), Type 1

IEC 60804 (2000-10), Type 1

DIN 45657 (1997-07)

IEC 61260 (1995) Octave and 1/3-octave Bands Class 1

ANSI S1.4 – 1983 plus ANSI S1.4A-1985 Amendment, Type 1

ANSI S1.43 – 1997, Type 1

Conforms with the following National and International Filter 
Standards:

IEC 61260 (1995-07) plus Amendment 1 (2001-09), 1/3-octave 
Bands, Class 0

ANSI S1.11-1986, 1/3-octave Bands, Order 3, Type 0-C

ANSI S1.11-2004, 1/3-octave Bands, Class 0

Note:

 The International IEC Standards are adopted as European 

standards by CENELEC. When this happens, the letters IEC are replaced 
with EN and the number is retained. Type 3639-G/3639-E also 
conforms to these EN Standards

SELF-GENERATED NOISE LEVEL

Typical A-weighted values at 23

°

C for nominal microphone sensitivity:

Type 3639-G (Microphone Type 4184)

Microphone:

 24.6 dB

Electrical:

 21.4 dB

Total:

 26.3 dB 

Type 3639-E (Microphone Type 4952)

Microphone:

 14.1 dB

Electrical:

 17.2 dB 

Total:

 18.9 dB

MEASURING RANGES (FOR SOUND LEVEL METER)
Dynamic Range:

 From typical noise floor to max. level for a 1 kHz pure 

tone signal, A-weighted:

Type 3639-G (Microphone Type 4184), Range:

 26.3 – 142.0 dB

Type 3639-E (Microphone Type 4952), Range:

 18.9 – 134.0 dB 

Linear Operating Range:

 In accordance with IEC 61672, A-weighted, 

1 kHz:

Type 3639-G (Microphone Type 4184), Range:

 34.1 –  142.0 dB

Type 3639-E (Microphone Type 4952)

Range:

 29.8 – 134.0 dB

Peak C Range:

 In accordance with IEC 61672:

Type 3639-G (Microphone Type 4184), Range:

 51.5 – 145.0 dB

Type 3639-E (Microphone Type 4952), Range:

 47.4 – 132.3 dB

MEASURING RANGES (FOR FREQUENCY ANALYZER)
Linear Operating Range:

 In accordance with IEC 61260:

Type 3639-G (Microphone Type 4184)

Range:

 

47.0 – 142.0 dB

Type 3639-E (Microphone Type 4952), Range:

 

48.3 – 129.3 dB 

CORRECTION FILTERS

For Microphone Types 4184 and 4952, Type 3639-G/3639-E is able to 
correct the frequency response to compensate for sound field (free-
field or diffuse-field) and for Type 4952 0

°

 (top) reference direction 

and 90

°

 (side) reference direction

The CE marking is the manufacturer's declaration that the product meets the requirements of the applicable EU directives
RCM mark indicates compliance with applicable ACMA technical standards – that is, for telecommunications, radio 
communications, EMC and EME
WEEE mark indicates compliance with the EU WEEE Directive

Safety

EN/IEC 61010–1 and ANSI/UL 61010-1: Safety requirements for electrical equipment for measurement, control and 
laboratory use.

EMC Emission

EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light industrial environments.
EN/IEC 61000–6–4: Generic emission standard for industrial environments.
CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits.

EMC Immunity

EN/IEC 61000–6–1: Generic standards – Immunity for residential, commercial and light industrial environments.
EN/IEC 61000–6–2: Generic standards – Immunity for industrial environments.
EN/IEC 61000–4–4: Fast transient/burst immunity. Level 2 (

±

1000 V)

EN/IEC 61000–4–5: Surge immunity. Line-to-line: Level 1 (

±

500 V). Line-to-ground: Level 3 (

±

2000 V)

Note:

 The above is guaranteed only: 

If Noise Monitoring Terminal Plus Type 3639-E is correctly assembled according to the instructions given in the User Manual
When using accessories listed in this Product Data

Temperature

IEC 60068–2–1 & IEC 60068–2–2: Environmental Testing. Cold and Dry Heat.
NMT Cabinet and Contents:
Operating Temperature: –10 to +50

°

C (14 to 122

°

F)

Operating Temperature with low-temperature protection kit: Down to –40

°

C (–40

°

F) 

Storage Temperature: –25 to +70

°

C (–13 to +158

°

F)

Humidity

IEC 60068–2–78: Damp Heat: 93% RH (non-condensing at 40

°

C (104

°

F)). Recovery time 2 to 4 hours

Mechanical

Non-operating:
IEC 60068–2–6: Vibration: 0.3 mm, 20 m/s

2

, 10 – 500 Hz

IEC 60068–2–27: Shock: 500 m/s

2

IEC 60068–2–29: Bump: 1000 bumps at 250 m/s

2

Enclosure

IEC 60529: Protection provided by enclosures: IP 55 (without external cables); IP 44 (with external cables)

Содержание 3639-E

Страница 1: ...ptional Weather data monitoring optional Camera support optional Remote operation via LAN public telephone lines mobile phone or GPRS Modular hardware software for easy on site servicing Noiseeventtri...

Страница 2: ...monitoring terminal and its accessories NMT PLUS Type 3639 E or NMT ADVANCED Type 3639 G Includes Wall Pole Mounting Kit UA 1685 Outdoor Microphone 4952 NMT 3639 E Weatherproof Microphone 4184 NMT 36...

Страница 3: ...noise monitoring software s server Operate as a broadband terminal with no noise event detection thus permitting a very cost effective NMT both with growth potential and with the possibility for remot...

Страница 4: ...el Kj r During transport the microphone can be safely placed inside the NMT cabinet For more information see Product Data BP 2099 NMT Microphone Central Control Software Communication Protocols Suppor...

Страница 5: ...n kit is available for continuous operation of the NMT under extreme weather conditions when running with mains power supply The NMT also includes a watchdog function that ensures that in the rare eve...

Страница 6: ...wer consumption The NMT complies with IEC 61672 Class 1 specifications and has a dynamic range of 110 dB Broadband Measurements and Frequency Analysis The NMT performs all the broadband analyses neede...

Страница 7: ...and liquid precipitation can be monitored Alternatively a two parameter Weather Station MM 0316 for windspeed and direction can be installed The external weather station mounted for example on a mast...

Страница 8: ...s are done using Noise Monitoring Software Type 7802 see Product Data BP 1573 The 300 series NMT configurations allow remote control of the NMT by customer defined third party software through the doc...

Страница 9: ...39 E Microphone Type 4952 Range 48 3 129 3 dB CORRECTION FILTERS For Microphone Types 4184 and 4952 Type 3639 G 3639 E is able to correct the frequency response to compensate for sound field free fiel...

Страница 10: ...nt on the size of the hard disk Standard configuration consists of a 40 GB hard disk giving at least six months storage space 90 days for Type 3639 E G 200 series Retention Time User defined INTERFACE...

Страница 11: ...ENMS and ANOMS including GPRS Router Type 3639 G 300 Noise Monitoring Terminal Advanced for System Integrators Type 3639 G 301 Noise Monitoring Terminal Advanced for System Integrators including PSTN...

Страница 12: ...hone 45 77 41 20 00 Fax 45 45 80 14 05 www bksv com info bksv com Local representatives and service organisations worldwide BP 2098 _ BP 2098 14 2014 06 Br el Kj r reserves the right to change specifi...

Отзывы: