4. TROUBLESHOOTING
4.1
Troubleshooting for Error Display
4.1.1 All scanned data are printed as blank page.
4.1.2 The scanned data has exceeded the memory capacity when using USB.
The number of files or folders has exceeded the upper limit when using USB.
4.1.3 Security communication failed.
4.1.4 Authentication by set user name and password failed.
Access request to the server timed out.
Transmission with "to FTP", "to SFTP", "to Network" or "to SharePoint" failed.
4.1.5 The scanned data has exceeded the memory capacity of the machine.
The scanned data has exceeded the memory capacity of the machine when "to
PC" was executed.
The scanned data has exceeded the memory capacity of the machine when
"WSScan" was executed.
Step
Cause
Remedy
1
The maintenance mode function
"Acquire white level data and set
CIS scan area" are not set.
Execute the maintenance mode function
"Acquire white level data and set CIS scan
area". (Refer to
2
Main PCB failure
Replace the main PCB ASSY.
<User Check>
- Perform scanning by dividing the document.
- Use an USB flash memory with more memory capacity.
- Change the USB flash memory.
Step
Cause
Remedy
1
Main PCB failure
Replace the main PCB ASSY.
<User Check>
- Check the password.
Step
Cause
Remedy
1
Main PCB failure
Replace the main PCB ASSY.
<User Check>
- Check the password settings.
Step
Cause
Remedy
1
Main PCB failure
Replace the main PCB ASSY.
<User Check>
- Perform scanning by dividing the document.
Step
Cause
Remedy
1
Main PCB failure
Replace the main PCB ASSY.
Содержание ImageCenter ADS-2400N
Страница 23: ...2 4 Confidential 2 2 Paper Feeding Fig 2 2 Feed path Front side Back side ...
Страница 46: ...3 2 Confidential 2 PACKING ...
Страница 49: ...3 5 Confidential 5 LUBRICATION There are no applicable parts for lubrication ...
Страница 50: ...3 6 Confidential 6 OVERVIEW OF GEARS There are no gears to be disassembled ...
Страница 96: ...4 3 Confidential 6 Click the Machine Info The firmware version is displayed ...