Host Commands
Polycold Cryochiller
Compressor Sensor Readings
Installation and Operation Manual
Brooks Automation
7-30
214072 Revision B
COMP_DISCHRG_TEMP Compressor Discharge Temp
COMP_LINE_TEMP Compressor Liquid Line Temperature
COMP_SUCT_PRESS Compressor Suction Pressure
Purpose:
Read the compressor discharge temperature sensor value. (TC1)
Parameter(s):
[n],?
[n] is
0 = Refrigerant compressor
1 = Not Used
Return Value:
[A][nnn.n]
Default Value:
N/A
Security Level
Get: 1
Set: N/A
Example:
“COMP_DISCHRG_TEMP,0,?”
Interface
RS-232, Ethernet
Purpose:
Read the compressor liquid line temperature sensor value. (TC2)
Parameter(s):
[n],?
[n] is
0 = Refrigerant compressor
1 = Not Used
Return Value:
[A][nnn.n]
Default Value:
N/A
Security Level
Get: 1
Set: N/A
Example:
“COMP_LINE_TEMP,0,?”
Interface
RS-232, Ethernet
Purpose:
Read the compressor suction pressure sensor value.
Parameter(s):
[n],?
[n] is
0 = Refrigerant compressor
1 = Not Used
Return Value:
[A][nnn.n]
Default Value:
N/A