Brooks On-Board IS Скачать руководство пользователя страница 1

4

Copyright 2013, Brooks Automation Inc

Part Number 8040648, Revision A, 01/11/2013

ECO Number 63723

Brooks Automation 

www.brooks.com

Printed in U.S.A

1

800-367-4887

 USA

978-262-2400

 International

Host RS-232 Connector

Table 1 

 provides information for connecting the Host computer to the On-Board 

IS

 Controller.

Table 1: Host RS-232 Connector Specifications

Parameter

Value

Baud Rate

9.6 kbs

Data Bits

7

Parity

Even

Number of Stop Bits

0

NOTE: 

The Host computer RS-232 cable must be fully shielded through to the outer shell. Use CTI-Cryogenics cable 

8132157 or equivalent.

Startup

See the 8040647, 

On-Board 

IS

 Cryopump Operation Instructions

, for details.

Status LEDs

Status LED III blinks when the On-Board 

IS

 Controller is operating, and indicates normal Intercomponent Network 

communication.Status LED I and LED II remain off during normal operation.

Helium Mapping

A helium map is a collection of cryopumps and compressors that the On-Board 

IS

 Controller manages, so that the 

system operates efficiently. The map controls when each cryopump uses a shared helium manifold and compressor for 
each process tool. See the 

On-Board IS Cryopump System Operation Guide

, CTI-Cryogenics part number 8040647, 

for more information about helium maps.

To check the device (cryopumps and compressors) addresses in a helium map:

1. Make a physical inventory of the system, noting the address of each device.

 

2. Use the Remote keypad (part number 8187007K00, see 

Figure 3

) to go to the 

On-Board 

IS

 Controller

 screen.

3. Choose 

Monitor 

and press 

Enter

, choose 

Show Devices 

and then press 

Enter

.

4. Ensure that all device addresses appear correctly.

 

If all device addresses match the physical inventory, continue to add or change a helium map. If the device addresses do not 

match, check the cable connections and repeat from 

Step 2

 through this step.

 

To change or add to a helium map:

1. Use the Remote keypad to go to the 

On-Board IS Controller

 screen. 

2. Choose 

System Setup 

and press 

Enter

, choose 

Helium

 and then press 

Enter

, finally choose 

Helium Map X

 

(where 

X

 is the number of the helium map you want to see) then press 

Enter

. The 

Choose Map Pumps 

screen appears.

3. Choose the addresses of cryopumps that match your inventory addresses, go to 

Accept Change

, and then 

press 

Enter

.

 

The 

Choose Compressors 

screen appears.

4. Choose the addresses of compressors that match your inventory addresses, go to 

Accept Change

, and then 

press 

Enter

.

 

The 

Verify Helium Map X 

screen appears.

5. Go to 

Accept Change

 and press 

Enter

 to set the helium map.

 

If you want to make changes to the helium map, press the back button as necessary and make the changes.

Product Information and Technical Support

Please visit the Brooks Automation website at www.brooks.com or email to 

[email protected].

On-Board

®

 

IS

 Controller (Pump Mount)

Quick Installation Guide

Part Number 8040648, Revision A, 01/11/2013

ECO Number 63723

On-Board 

IS

 Controller Specifications

Input Voltage and Power: 24 VDC +/- 10% (21.6 - 26.4) VDC - 18 watts

Ambient Temperature: 50º F - 100º F (10º C - 38º C)

Communication Interface: Host RS-232, Aux RS-232, Service RS-232 - 9 Pin D Connector

Communication Interface Baud Rate: 9600

On-Board 

IS

 Intercomponent Network Interface: RS-232, RS-485

On-Board 

IS

 Remote Interface: USB B Connector with RS-232 Protocol