background image

CoolBox

TM

 30

Instructions for Use

Overview of CoolBox 30 System
CoolBox 30 is a portable cooling system for maintaining cooling and freezing sample temperature 
on the bench top without ice. It is intended for use with a variety of CoolRack

®

 tube modules. 

The design and thermal conductivity of CoolRack modules maintain samples at uniform well 
to well temperature throughout the cooling period regardless of sample position. By selecting 
from a number of cartridges and cooling options, samples can be maintained within a variety of 
temperature bands that range from -78°C to 4°C.  For a list of compatible CoolRack modules, see 
last page. 

Using +2

o

C Gel Cartridge (Blue) at  0.5

o

C to 4

o

C  

Freeze blue +2°C cartridges in a -20°C freezer for a minimum of four hours. Pre-chill the CoolRack 
module to 0 to 4°C in a refrigerator for approximately 30 minutes. Remove cartridge from freezer 
and allow cartridge to sit at room temperature until the surface frost begins to melt (approximately 
10 minutes). Place a frozen cartridge in the CoolBox 30 base followed by a CoolRack module. The 
CoolRack temperature should equilibrate between 0.5°C to 4°C. Keep the lid on the CoolBox 30 to 
prolong the cooling duration. 

IMPORTANT: It is possible to also start with a room temperature CoolRack module placed directly 
onto frozen cartridge immediately after removing it from the freezer. The system will take approxi-
mately 20 minutes to equilibrate to approximately 4°C and will not undercool samples below 0°C. 

For cooling beyond 10 hours, simply exchange the cooling cartridge with a freshly charged one.

Actual performance may vary depending upon ambient temperature, start-up parameters, sample load, 

initial sample temperature, air currents, radiant energy sources and other conditions. 

Temperature Range

Cooling Source

Cooling Duration

Open Lid

Cooling Duration

Closed Lid

0.5

o

C to 4

o

C

+2

o

C Cartridge - Blue

Over 4 hours

Over 10 hours

-20

o

C to -4

o

C

-20

o

C to 0˚C Cartridge - Green

Over 3 hours

Over 6 hours

-78

o

Dry Ice

Over 5 hours

Over 6 hours

Care and Cleaning
CoolBox 30 is constructed from high density closed cell polyethylene foam. The CoolBox 30 is com-
patible with prolonged cryogenic temperature exposure. The foam may be cleaned by water and 
mild soap. Rinse thoroughly. The CoolBox 30 is resistant to alcohols and 10% bleach solutions. 
Avoid abrasive or sharp objects. Do not use the CoolBox for pulverizing dry ice. Do not autoclave. 
Maximum temperature exposure: 60°C. Avoid exposure to UV light sources. The CoolBox 30 is not 
recommended for shipping of temperature-sensitive samples.

CoolRack and CoolSink Thermo-Conductive Sample Modules Compatible with CoolBox 30

Dimensions (L x W x H):
Internal (with Cooling Cartridge loaded): 12 x 10.4 x 7.9 cm / 4.7 x 4 x 3.1 in   
External: 16.5 x 15.25 x 12.7 cm / 6.5 x 6 x 5 in 

Care and Cleaning

The CoolBox is constructed from high density polyethylene foam.   The CoolBox is compatible with prolonged cryogenic tempera-
ture exposure.  The CoolBox may be cleaned by water and mild soap.  Rinse thoroughly.  The CoolBox is resistant to alcohols and 
10% bleach solutions.   Avoid abrasive or sharp objects.  Do not use the CoolBox for pulverizing dry ice.   Do not autoclave.  Maxi-
mum temperature exposure: 60°C.  Avoid exposure to UV light sources.  

0.5°C to 4°C (+2ºC Gel Cartridge - Blue)

Freeze blue +2°C cartridges in a -20°C freezer for a minimum of four hours.  Pre-chill the CoolRack to 0 to 4°C in a refrigerator 
for approximately 30 minutes. Remove cartridge from freezer and allow cartridge to sit at room temperature until the surface 
frost begins to melt (approximately 10 minutes).  Place a frozen cartridge in the CoolBox followed by a CoolRack. The CoolRack 
temperature should equilibrate to between 0.5°C to 4°C.  Keep the lid on the CoolBox to prolong the cooling duration; use 
insulator pad to extend cartridge service interval or in conditions of higher environmental temperatures.

Important: It is possible to also start with a room temperature CoolRack placed directly onto a frozen cartridge immediately after 
removing it from the freezer.  The system will take approximately 20 
minutes to equilibrate to approximately 4°C and will not undercool 
samples below 0°C. 

For cooling beyond the specified times shown below simply exchange 
the cooling cartridge with a recharged one.

Use of insulator pad for extended cooling time:
Using the insulator pad on top of the CoolRack or closing the lid will 
significantly enhance cooling time.  Tubes can be inserted through the 
pad.

Temperature Measurement and CoolBox Validation

The CoolRack thermoconductive design ensures that all well temperatures are identical.  

Test equipment:
Temperature probe with small thermal mass such as a K-type beaded thermocouple probe. 
Digital thermometer to match probe.
Representative closed sample tube with cap

Test procedure:
Drill a small hole in the center of the sample tube cap with a diameter that provides a snug fit for the temperature probe wire so 
there will be minimal ambient air influx into the tube. 
Fill the sample tube with 1ml of water. 
Place the cap on sample tube and insert temperature probe. 
Place the instrumented sample tube in any well of the CoolRack.  
Rack temperatures can be measured by direct contact of the probe in any of the rack wells.

Optional Insulator

CoolRack PF30,M30,CF30

CoolBox

+2ºC Cartridge

The CoolBox

 30 is a portable device for maintaining cooling and freezing sample temperatures on the bench top.  It is intended for 

use with any of the 30-well CoolRack

 tube holders such as CoolRack M30, M30-PF, CF30, CFT30, 500ul and MassSpec.  Due to the 

design and thermal conductivity of CoolRacks, samples will experience identical well to well temperature throughout the cooling 
period regardless of sample position.  By selecting from a number of cartridges and cooling options, samples can be maintained within 
a variety of temperature bands that range from -78°C to 4°C.

CoolBox

Temperature Range

Cooling Source

Cooling Duration

Open Lid 

Cooling Duration

Closed Lid

0.5ºC to 4ºC

+2ºC Cartridge - Blue

up to 4 hours

up to 10 hours

0.5ºC to 4ºC

Wet Ice Cartridge - Red

up to 4 hours

up to 10 hours

-18ºC to -4ºC

-12ºC Cartridge - Green

up to 3 hours

up to 6 hours

-30ºC to -20ºC

Dry Ice with 

Insulator Pad

up to 5 hours

up to 10 hours

-78ºC

Dry Ice

up to 5 hours

  up to 6 hours

copyright 2008, BioCision

®

, LLC.  All rights reserved.  Patents pending.

CoolBox Instructions, rev October 2010

www.biocision.com

775 E. Blithedale, Suite 203    Mill Valley, CA  94941    Tel 888.478.2221    Fax: 415.634.2350    Email:[email protected]

Note 1:  Using insulator pad to protect exposed surface of CoolRack from environmental heat influx will provide optimal cooling duration.
Note 2:  Actual performance may vary depending upon ambient temperature, start-up parameters, sample load, initial sample temperature, 
air currents, radiant energy sources and other conditions.

0

4

10º 

3

10 hours

Lid 

Lid 

* Data obtained from pre-chilled (4°C) CoolRack at 
approximately 18-20°C ambient room temperature.

*Data obtained from pre-chilled (4˚C) CoolRack at 
approximately 18- 20˚C ambient room temperature.

Base

CoolRack module

Cooling Cartridge 

(Pre-frozen in -20˚C 

Freezer)

Lid

Item

Description

Accommodates

BCS-125

CoolRack M15

15 x 1.5ml or 2ml microfuge tubes

BCS-127

CoolRack M15-PF

15 x 1.5ml conical microfuge tubes

BCS-108

CoolRack M30

30 x 1.5ml or  2.0ml microfuge tubes

BCS-128

CoolRack M30-PF

30 x 1.5ml conical microfuge tubes

BCS-137

CoolRack M-PF 500 uL

30 x 500 uL microfuge tubes

BCS-126

CoolRack CF15

for 15 cryogenic vials

BCS-138

CoolRack CFT30

for 30 cryogenic vials or FACS tubes

CAUTION: The product described here are intended for the exclusive use by trained and experienced 

laboratory and medical personnel. A portion of the methods described require the use of dry ice. Direct 
skin contact with dry ice or metal components that have been touching dry ice can cause freezing injury. 
Always use appropriate protective equiptment for eyes and skin when handling dry ice and cold metal 
components. 

Made in USA.  

15 Elizabeth Dr., Chelmsford, MA 01824 USA  brookslifesciences.com/contact-us  brookslifesciences.com

CoolBox 30 Instructions, May 2018

CoolBox 30 ©Brooks Automation, Inc. All rights reserved.

Отзывы: