Fieldbus 2 Baud Rate
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned long
RW
B
0…1.0E10
310
124/9
0xFC48…0xFC49/64585…64586
Fieldbus 2 Parity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
B
0…2
336
124/12
0x0F8C/3981
The following values are supported:
Value
Description
0
No parity
1
Odd parity
2
Even parity
4.5
Fluid set
Fluid Set Index
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…7
24
1/16
0x0030/49
With this parameter, any of the pre-configured fluids (up to 8) can be selected. Each fluid has its specific (configurable)
properties, such as
Fluid Name
,
Capacity
, etc.
Default value: 0 (fluid 1).
Note that the selected value is equal to the fluid number minus 1 (value 0 corresponds to fluid 1, value 1 to fluid 2, etc.)
Fluid Name
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[10]
RW
B
-
25
1/17
0x8188…0x818C/33161…33165
This parameter contains the name of the current fluid.
Capacity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
B
1E-10…
1E+10
21
1/13
0x8168…0x8169/33129…33130
This parameter sets the maximum readout/control value (100%) for the current fluid in readout units corresponding to
Capacity Unit
.
Capacity Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[7]
RW
B
see below
129
1/31
0x81F8…0x81FB/33273…33276