16
2.6.3.2
Modbus
Modbus is a 3-wire, RS-485-based fieldbus communication system for parameter value exchange. In this system each
instrument/device is equipped with a micro-controller for its own dedicated task. The instrument behaves as a slave, which
means all communication (instructions and readout) is initiated by a master device on the Modbus system.
Characteristics:
·
Baud rate selectable between 9600 and 256000 Baud (default: 19200 Baud)
·
+15…24 Vdc supply voltage
·
Connection of up to 247 instruments on a single bus
·
Supports RTU and ASCII protocols
Consult
Instruction manual Modbus interface
(document no. 9.17.035) for more information about setting up a
Modbus network.
Detailed information about Modbus can be found at
or any website of the (local) Modbus
organization of your country (if available).
2.6.3.3
PROFIBUS DP
PROFIBUS DP is a 2-wire, RS-485-based industrial data communication standard (fieldbus) which allows automation
components (e.g. sensors, actuators and controllers) to exchange information.
Consult
Instruction manual PROFIBUS DP interface
(document no. 9.17.025) for more information about setting up a
PROFIBUS DP network.
2.6.3.4
DeviceNet™
The DeviceNet™ interface offers a direct connection to a DeviceNet™ network, according to the mass flow controller profile
specified by the ODVA. The Bronkhorst® DeviceNet™ instrument is a Group 2 Only Server device whose messages comply
with the Controlled Area Network (CAN) 2.0A standard and with the DeviceNet™ protocol.
Consult
Instruction manual DeviceNet™ interface
(document no. 9.17.026) for more information about setting up a
DeviceNet™ network.