Bronkhorst®
Instruction Manual FLEXI-FLOW™ Compact
9.17.158A
28
Default settings
Network configuration is done ex factory as ordered. The table below shows the supported configurations for the available
interface protocols (default settings are printed in bold):
Protocol
FLOW-BUS
Modbus RTU
Address
3
…125
1
…247
Baud Rate
187500
9600
19200
38400
56000
57600
115200
128000
256000
Parity
0
0, 1,
2
acts as an individual instrument, all three communication channels of the
FLEXI-FLOW™ are accessible on the same node address.
4.2.5
Fluid set
Fluid Set Index
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…7
24
1/16
0x0030/49
With this parameter, any of the pre-configured fluids (up to 8) can be selected. Each fluid has its specific (configurable)
properties, such as
Fluid Name
,
Capacity
, etc.
Default value: 0 (fluid 1).
Note that the selected value is equal to the fluid number minus 1 (value 0 corresponds to fluid 1, value 1 to fluid 2, etc.)
Fluid Name
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[10]
RW
B
-
25
1/17
0x8188…0x818C/33161…33165
This parameter contains the name of the selected fluid.
Capacity 100%
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
B
1E-10…
1E+10
21
1/13
0x8168…0x8169/33129…33130
·
This parameter represents the 100 % readout/control value (span) for the selected fluid, expressed in the selected
Capacity
Unit
.
·
Capacity 100%
is scaled when
Inlet Pressure
,
Fluid Temperature
or
Fluid Name
is changed for the selected fluid.
Capacity Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[7]
RW
B
see below
129
1/31
0x81F8…0x81FB/33273…33276
15