Bronkhorst®
Instruction Manual ES-FLOW™ 1xxC
9.17.153A
28
Counter Unit
supports the following values:
Mass
Normal volume
(1.01325 bar(a), 0 °C)
Standard volume
(1.01325 bar(a), 20 °C)
Custom volume
(
Capacity Unit Pressure
,
Capacity Unit Type
Temperature
)
ug, mg, g, kg
uln, mln, ln,
mm3n, cm3n, dm3n, m3n
uls, mls, ls,
mm3s, cm3s, dm3s, m3s
ul, ml, l,
mm3, cm3, dm3, m3
Counter Value
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…
10000000
122
104/1
0xE808…0xE809/59401…59402
Current counter value in units selected with parameter
Counter Unit
.
Counter Limit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…9999999
124
104/3
0xE818…0xE819/59417…59418
Counter limit/batch size in units selected with parameter
Counter Unit
.
Default value: 0.
Counter Setpoint Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…1
126
104/5
0x0D05/3334
Specifies whether or not to change the setpoint after reaching the counter limit.
Value
Description
0
No setpoint change (default)
1
Change setpoint to
Counter new setpoint
Counter New Setpoint
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
RW
0…32000
127
104/6
0x0D06/3335
New (safe) setpoint when a counter limit is reached until reset. Range 0…32000 represents 0…100% setpoint.
Default value: 0
Reset Counter Enable
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…15
157
104/9
0x0D09/3338
Available reset methods for counters. Up to 3 different methods can be specified. The value is a bitwise summation of the
enabled reset methods; convert the value to binary to see which methods are enabled.
Default value: 7 (bits/methods 0, 1 and 2 enabled)
The following methods are supported:
Bit
Value
Description
0
1
Multifunctional switch
1
2
Externally
2
4
By parameter
Reset
3
8
Automatically (e.g. when counter value is reset)