Bronkhorst®
Instruction Manual EL-FLOW® Prestige
9.17.084N
41
Capacity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
1E-10…
1E+10
21
1/13
0x8168…0x8169/33129…33130
This parameter sets the maximum readout/control value (100%) for the current fluid in readout units corresponding to
Capacity Unit
.
Capacity Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[7]
RW
see below
129
1/31
0x81F8…0x81FB/33273…33276
Available units:
Mass flow
Normal volume flow
(1.01325 bar(a), 0 °C)
Standard volume flow
(1.01325 bar(a), 20 °C)
Custom volume flow
(
Capacity Unit Type Pressure
,
Capacity Unit Type
Temperature
)
ug/h, ug/min, ug/s,
mg/h, mg/min, mg/s,
g/h, g/min, g/s,
kg/h, kg/min, kg/s
uln/h, uln/min, uln/s,
mln/h, mln/min, mln/s,
ln/h, ln/min, ln/s,
ccn/h, ccn/min, ccn/s,
mm3n/h, mm3n/m,
mm3n/s,
cm3n/h, cm3n/m, cm3n/s,
m3n/h, m3n/min, m3n/s,
scfh, scfm, scfs,
sccm, slm
uls/h, uls/min, uls/s,
mls/h, mls/min, mls/s,
ls/h, ls/min, ls/s,
ccs/h, ccs/min, ccs/s,
mm3s/h, mm3s/m, mm3s/s,
cm3s/h, cm3s/m, cm3s/s,
m3s/h, m3s/min, m3s/s
ul/h, ul/min, ul/s,
ml/h, ml/min, ml/s,
l/h, l/min, l/s,
cc/h, cc/min, cc/s,
mm3/h, mm3/m, mm3/s,
cm3/h, cm3/m, cm3/s,
m3/h, m3/min, m3/s,
cfh, cfm, cfs
Because of the maximum string length (7 characters), some unit names are abbreviated. For instance mm3n/m means
mm
3
n/min.
Capacity Unit Type Temperature
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
-273.15…
3.4E+38
245
33/10
0xA150…0xA151/41297…41298
This parameter defines a reference temperature for conversion of the measured mass flow to a volume flow. See also
parameters
Capacity Unit
and
Counter Unit
.
Capacity Unit Type Pressure
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…3.4E+38
246
33/11
0xA158…0xA159/41305…41306
This parameter defines a reference pressure for conversion of the measured mass flow to a volume flow. See also
parameters
Capacity Unit
and
Counter Unit
.
Controller Speed
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0.2…5
254
114/30
0xF2F0…0xF2F1/62193…62194
This parameter sets the overall controller speed factor for the selected fluid set.
Controller Speed
is set ex factory between
value '0.5' (slow) and '2' (fast). The default value is '1'. Slower or faster settings are possible between values 0.2 and 5 at the
customer's responsibility.