Brentwood IRT-901 Thermometer Quick Start Guide
Instructions for Measuring Body Temperature
Using a non-contact infrared thermometer (NCIT) is an effective and fast method of measuring body temperature
but does require a little practice to ensure the best results. Follow the steps below.
NOTE: This is not a treatment, merely a measurement. Each personal temperature can vary daily and from person
to person. Use the memory setting to track your daily temperature. Always consult your doctor if you have
questions.
When turning on your device for the first time:
Press the “MODE” button for 2 seconds, the screen will display “F1”. Press the “MODE” button to transfer between
Celsius or Fahrenheit. Press the “MEM” button to confirm.
This will take you to “F2,” the Warning Adjustment Screen, where you can set the temperature at which the “HI”
error appears.
If the temperature taken is higher than the selected temperature
in this setting, the “HI” error will appear.
The default temperature at which the “HI” error appears is
100.4°F
.
You can adjust down by pressing the “MODE” button.
It will go as low as 99.1°F and as high as 102.4°F before it
resets to the default.
Once you have selected the “HI” error temperature, press the
“MEM” button.
This will take you to “F4,” the Recalibration Screen, where you
can set the temperature offset.
Press the “MODE” button to choose the Temperature Offset
value from -5.4°F to 5.4°F.