background image

PHOTODYNAMIC THERAPY

HELBO

®

User Manual   

HELBO®

 

TheraLite Laser

 

and additional accessories for the antimicrobial Photodynamic therapy

Version 14  06/20

17

bredent medical GmbH 

&

 Co. KG 

|

 Weissenhorner Str. 2 

|

 89250 Senden 

|

 Germany 

|

 T: (+49) 0 73 09 / 8 72-4 41 

|

 F: (+49) 0 73 09 / 8 72-4 44 

|

 www.bredent-medical.com 

|

 @: [email protected]

S  I  N  C  E     1  9  7  4

Содержание HELBO TheraLite

Страница 1: ...additional accessories for the antimicrobial Photodynamic therapy Version 14 06 2017 bredent medical GmbH Co KG Weissenhorner Str 2 89250 Senden Germany T 49 0 73 09 8 72 4 41 F 49 0 73 09 8 72 4 44 w...

Страница 2: ...D Pocket Probe 13 2 4 2 HELBO 2D Spot Probe 14 2 4 3 HELBO 3D Endo Probe 15 2 5 START UP 16 2 5 1 Batteries 16 2 5 2 Operating states displays 17 2 5 3 Main switch 18 2 5 4 Button 18 2 6 MAINTENANCE S...

Страница 3: ...ls Staining the pathogenic bacteria with the dye and then exposing them to the laser results in targeted transfer of the light energy to the ambient oxygen converting it quantum energetically into sho...

Страница 4: ...sional uniform lateral exposure Increased apical radiation More flexible fibre optic for effective use in root canals Greater probe rigidity Sterile disposable product to prevent spread of bacteria Po...

Страница 5: ...Rinsing using the multi functional syringe and suctioning off excess dye Rinsing and suction of excess phenothiazine chloride dye None Photosensitizer display blue flashes rapidly None Rinsing using...

Страница 6: ...ill prevent the phenothiazine chloride dye from reaching all the bacteria 3 Leave the phenothiazine chloride dye on for at least one minute three minutes when treating pockets 6 mm or persistent refra...

Страница 7: ...treating small areas ISO45 use a suitable instrument e g Lentulo to transfer the dye apically 4 Leave the phenothiazine chloride dye for at least one minute Important Only bacteria stained blue will...

Страница 8: ...upply are reduced Why does the photobiological effect work The radiation with blue to red laser light wavelength 380 nm to 700 nm stimulates the synthesis from ADP to ATP through oxidative phosphoryla...

Страница 9: ...r is a battery powered hand held mobile Class 2M laser device The front part of the HELBO TheraLite Laser has a fibre optic connector Three different coloured LEDs near the button indicate the device...

Страница 10: ...nd each have specific indications The treating physician or dentist is solely responsible for selecting patients to receive this treatment Potential dental indications include Periodontal disease Peri...

Страница 11: ...dam or coating them with petroleum jelly prior to application If dye is applied directly to fillings or crowns rinse it away using a great deal of water Discolouration of surrounding tissue should di...

Страница 12: ...HELBO 3D Endo Probe and HELBO 2D Spot Probe are all shipped in sterile packaging as sterile ready to use products Remove them from the packaging under sterile conditions To do this grasp the silver c...

Страница 13: ...r bring the HELBO 3D Pocket Probe in contact with infected tissue regions as much as possible Make sure any pockets or recessed areas within the tissue receive full exposure When treating areas with c...

Страница 14: ...d also serves as orientation for overlapping exposure The HELBO 2D Spot Probe is particularly suitable for the following indications Periodontal infections treated by open curettage Root tip resection...

Страница 15: ...ours even light distribution across the entire lateral radiation area of 12 mm The HELBO 3D Endo Probe emits radiation with increased power in a forwards direction to ensure the apex is ade quately ir...

Страница 16: ...ies and processes required for antimicrobial Photodynamic therapy Brief overview of steps in the commissioning process 1 Insert batteries 2 Check operational status display 3 Turn main switch on 4 Pre...

Страница 17: ...ion display is located in front of the start button and can have the following states GREEN The combination display lights up green as soon as the batteries have been correctly inserted and the main s...

Страница 18: ...o turn off the HELBO TheraLite Laser After treatment turn the main switch to the OFF position and place the HELBO TheraLite Laserback into its case or into the HELBO T Controller Please recharge flat...

Страница 19: ...49 0 73 09 872 6 00 Telefax 49 0 73 09 872 6 35 2 7 CLEANING IMPORTANT Never autoclave the HELBO TheraLite Laser or immerse it in liquid Clean the HELBO TheraLite Laser using only a pH neutral disinfe...

Страница 20: ...s size AAA R03 with new freshly charged batteries approx 120 min in continuous operation The following chargers have been inspected for charging the aforementioned batteries Voltcraft IPC Sanyo NC MQR...

Страница 21: ...the batteries will then fall out of the battery compartment and insert new batteries recharge empty batteries immediately the positive terminal of the batteries must face in the direction of the main...

Страница 22: ...is not permitted Take precautionary measures to ensure that the laser beam does not hit surfaces that are reflective in the same way as a mirror particularly flat or concave reflective surfaces It is...

Страница 23: ...n once Displacement of optical fibre as a result of improper attempts to attach Defects in the HELBO 2D Spot Probe HELBO 3D Pocket Probe HELBO 3D Endo Probe can be recognised by the following signs Th...

Страница 24: ...ion The HELBO TheraLite Laser is intended for use in an environment such as that described below The customer or user of the HELBO TheraLite Laser must ensure that it is operated in such an environmen...

Страница 25: ...y frequency should be at levels typical for business and hospital environments Note UT is the AC mains voltage prior to application of the test level Guidelines and manufacturer s declaration Electrom...

Страница 26: ...reflection from buildings objects and people a The field strength of stationary transmitters such as mobile phone and land based mobile radio base stations amateur stations AM and FM radio and televis...

Страница 27: ...e as indicated below Nominal output of transmitter W Safety margin dependent on transmission frequency m 150 kHz to 80 MHz P V d 1 5 3 80 MHz to 800 MHz P E d 1 5 3 800 MHz to 2 5 GHz P E d 1 7 0 01 0...

Страница 28: ...gns information signs and general signs Inspection of instruction manual available complete Inspection of locking device Calibration of laser output Measurement of actual laser output with calibrated...

Страница 29: ...A at 25 C Environmental conditions for storage Temperature 10 C to 40 C Relative humidity 30 to 95 Air pressure 700 hPa to 1060 hPa Environmental conditions for operation Temperature 10 C to 35 C Rela...

Страница 30: ...that the HELBO TheraLite Laser conforms to the specifications of EU Directive 93 42 EEC on medical devices The HELBO TheraLite Laser is manufactured in conformance with a certified quality assurance...

Страница 31: ...ent medical GmbH Co KG Weissenhorner Str 2 89250 Senden Germany Telefon 49 0 73 09 872 6 00 Telefax 49 0 73 09 872 6 35 2 15 TROUBLESHOOTING Problem Causes and solution Even when the batteries are ins...

Страница 32: ...e fibre optics lower intensity lighter beam Fibre optics not correctly attached Fibre optics damaged see Point Fibre Optics Combination display lights up green and or ange simultaneously Electronics e...

Страница 33: ...properly or is missing In cases of emergency removing the applicator will halt laser beam emission In the event of an emergency remove the fibre optics to stop laser emission If necessary the opening...

Страница 34: ...13 Year of manufacture 77 HELBO type index 0001 Sequential number See user manual Manufacturer DC supply voltage 2 x 1 2 volts Direction for inserting batteries Degree of protection against electrical...

Страница 35: ...f relative humidity for storage purposes 700hPa 1060hPa Details of air pressure for storage purposes 2 18 GENERAL INFORMATION ABOUT LASER SAFETY Low intensity therapy laser devices can be harmful to t...

Страница 36: ...with CE identification CE mark and Notified Body Number e g CE 0483 are to be sold Labelling only with the letters CE is not sufficient As a general rule anyone working or intending to work with low i...

Страница 37: ...YNAMIC THERAPY HELBO User Manual for HELBO TheraLite Laser 37 Version 14 06 2017 Mistake and subject to change reserved 009892GB 20170427 40 Good luck with aPDT and the HELBO TheraLite Laser Your HELB...

Отзывы: