7.3.7
Rebreathing Alarm
Property
Description
Alarm text
Rebreathing
Priority
High
Alarm condition
A Rebreathing alarm will be given if the intentional leakage is
too low for more than 15 seconds.
Possible cause
•
Obstructed or occluded patient circuit.
•
Incorrect patient circuit.
Reset criteria
The leakage is back within limits.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
On
•
Off
7.3.8
Apnea Alarm
Property
Description
Alarm text
Apnea
Priority
High
Alarm condition
An Apnea alarm will be given when no patient-triggered
breath is detected for the set period of time.
Possible cause
•
Patient stopped breathing.
•
Patient decreases spontaneous breathing.
•
Circuit disconnection.
•
Inspiratory Trigger is set too high.
Reset criteria
Inspiratory effort detected by the Vivo 3.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
5 to 60 s.
•
Off
Setting resolution
5 s below 15 s.
15 s above 15 s.
60 s above 60 s.
Alarms 99
Doc. 007232 Y-2