7.3.5
High Minute Volume Alarm
Property
Description
Alarm text
High MV
Priority
Medium
Alarm condition
A High Minute Volume alarm will be given when the
minute volume is above the set alarm limit for 15 seconds.
Possible cause
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the
patient circuit.
•
Increased breath rate.
Reset criteria
The minute volume goes below the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
1 l. to 40 l.
•
Off
Setting resolution
0.5 l.
7.3.6
Low Minute Volume Alarm
Property
Description
Alarm text
Low MV
Priority
High
Alarm condition
A Low Minute Volume alarm will be given when the minute
volume is below the set alarm limit for 15 1 seconds.
Possible cause
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the
patient circuit.
•
Decreased breath rate.
Reset criteria
The minute volume goes above the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
1 l. to 30 l.
•
Off
Setting resolution
0.5 l.
98 Alarms
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2