7.3.2
Low Pressure Alarm
Property
Description
Alarm text
Low Pressure
Priority
High
Alarm condition
A Low Pressure alarm will be given when the Vivo 3 pressure
fails to reach the low pressure alarm limit for 15 seconds.
Possible cause
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the
patient circuit.
Reset criteria
The pressure rises above the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
1 cmH
2
O to 40 cmH
2
O
•
Off
Note that the Low pressure alarm cannot be set higher than
the value set for IPAP or the High pressure alarm.
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Setting display
The Low Pressure alarm setting is displayed by a red line in
the pressure bar graph.
96 Alarms
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2