When finished with the sound level adjustment, check that the alarm can be received
at the operator
’
s position, see 7.1
7.3
Physiological Alarms
7.3.1
High Pressure Alarm
Property
Description
Alarm text
High Pressure
Priority
High
Alarm condition
A High Pressure alarm will be given when the patient pres-
sure reaches the set High Pressure alarm limit for three con-
secutive breaths.
Possible cause
•
Mismatch between pressure setting and alarm setting.
•
Coughing during inspiration.
•
Changes in airway resistance and or compliance.
Reset criteria
A full breath is performed with maximum pressure below the
alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings. The actual breath is however terminated if the High
Pressure alarm limit is reached.
Setting range
•
5 cmH
2
O to 45 cmH
2
O
•
Off
Note that the High pressure alarm cannot be set lower than
the value set for the Low pressure alarm.
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Setting Display
The High Pressure alarm setting is displayed by a red line in
the pressure bar graph.
Alarms 95