7.3.12
High SpO
2
Alarm
Property
Description
Alarm text
High SpO2
Priority
Medium
Alarm condition
A High SpO
2
alarm will be given when the measured SpO
2
exceeds the alarm limit for 30 seconds.
Possible cause
Too high flow of bleed-in oxygen.
Reset criteria
The SpO
2
value goes back below the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
80 % to 100 %
•
90 % to 100 %
•
Off
Setting resolution
1 %
This alarm requires a connected SpO
2
sensor.
7.3.13
Low SpO
2
Alarm
Property
Description
Alarm text
Low SpO2
Priority
High
Definition
A Low SpO
2
alarm will be given when the measured SpO
2
is
below the alarm limit for 30 seconds.
Possible cause
•
Too low flow of bleed-in oxygen.
•
Oxygen inlet is disconnected.
•
Delivered tidal volumes are too small.
Setting range
70% to 100%
Setting resolution
1%
Ventilator action
The ventilator will continue treatment with the same settings.
This alarm requires a connected SpO
2
sensor.
102 Alarms
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2