PV 101+/PV 102+ Service Manual
FAULT TRACING
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
8–1
8
FAULT TRACING
This chapter contains a fault-tracing table that may be used when trouble-shooting the
device.
8.1
Fault-tracing table
If the PV 101+/PV 102+ does not work properly try to identify the problem in the table
below. Check the possible causes and carry out the suggested remedial actions.
Symptom
Possible cause
Remedial action
See ref.
The device alarms at
start-up.
• The power cord is not pro-
perly connected.
• Connect the power cord.
• Mains fuse has blown.
• Replace the fuse.
The ventilator does
not give adequate
pressure/volume.
• Leakage from patient circuit/
mask.
• Check the tubes and mask
for leakage.
• Internal leakage.
• Check the tubes inside the
machine.
• Dirty filters.
• Replace or wash the filters.
• Pressure sensor faulty.
1 Adjust voltage to trans-
former.
2 Replace MDA board or
pressure sensor.
The PV 101+/
PV 102+ does not
trigger.
• Trigger settings not correct.
• Adjust settings.
See
PV 101+/
PV 102+
Operating
manual
• Tube to flow sensor has
worked loose.
• Check tube and connectors.
• Flow sensor faulty.
• MDA board faulty.
1 Adjust voltage to pressure
sensor.
2 Replace MDA board or flow
sensor.
Alarms for mask-off
(PV 102+ only).
• Leakage from patient circuit/
mask.
• Check the tubes and mask
for leakage.
Содержание PV 101+
Страница 28: ...CALIBRATION INSTRUCTIONS PV 101 PV 102 Service Manual 4 6 BREAS MEDICAL Doc No 003259 En Issue X 1...
Страница 76: ...FAULT TRACING PV 101 PV 102 Service Manual 8 2 BREAS MEDICAL Doc No 003259 En Issue X 1...
Страница 78: ...9 2 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Страница 82: ...9 6 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Страница 87: ...BREAS MEDICAL 10 PV 101 OPERATING MANUAL...
Страница 88: ...BREAS MEDICAL...
Страница 89: ...BREAS MEDICAL 11 PV 102 OPERATING MANUAL...
Страница 90: ...BREAS MEDICAL...